Case Report

-----------------------------------------------------------------------------------------------------

Inferior Vena Cava Filter Fracture and Migration: A Case Report and Review of the Literature

Kylie Lepic MD,FRCPC, JeffreyS. Ginsberg MD,FRCPC, Sonia S. Anand MD,PhD and Ameen Patel MB,FRCPC


About the Authors

Dr. Kylie Lepic received her MD from McMaster University in 2006. She completed her Internal Medicine and Hematology Residency training at McMaster followed by a Fellowship in Blood and Marrow Transplant at the Fred Hutchinson Cancer Research Center in Seattle,WA. Dr. Lepic is currently an assistant professor in the Department of Medicine at McMaster. Correspondence may be directed to lepic@hhsc.ca.

Abstract

Inferior vena cava (IVC) filters are an option for the prevention of pulmonary emboli in patients who have venous thrombosis and a contraindication to anticoagulation, but their placement is not without risk. We describe a case of a patient with a Gunther-Tulip IVC filter that fractured and migrated to the left distal pulmonary artery. The leg was removed using bronchoscopy forceps and a French sheath without complication. Fracture and migration ofiVC filters is rare with only a few case reports in the literature. it is important to consider given the potential for significant morbidity and mortality.

Resume

La mise en place de filtres dans la veine cave inferieure(VCI) constitue une option pour la prevention d'une embolie pulmonaire chez les patients qui presentent une thrombose veineuse et pour qui !'anticoagulation est contre-indiquee, mais cela n'est pas depourvu de risques. Nous decrivons le cas d'un patient dont le filtre Gunther-Tulip pour VCI s'est rompu et est aile se

loger dans l'artere pulmonaire distale gauche. La jambe a ete degagee a l'aide d'un forceps

bronchoscopique et d'une gaine French, et ce, sans complication. La rupture et le deplacement de filtres pour VCI est un phenomene rare;seuls quelques cas sont rapportes dans la revue de la litterature.11 est toutefois important de considerer cette eventualite compte tenu des risques de morbidite et de mortalite.

36 Vo lu me 11, Is su e 3, 2016 Ca n adia n Journa l of Ge n e r a l I n te r nal Medicine

Lepic et al.

Case Presentation

Inferior vena cava (IVC) filters are used for the prevention of pulmonary emboli in patients who have a contraindication to anticoagulation, but their placement is not without risk. Complications of IVC filters include hematoma at the insertion site, deep vein thrombosis (DVT), inferior vena cava thrombosis, postphlebitic syndrome, filter penetration into adjacent structures, filter fracture and fllter migration.Filter migration can be life threatening depending on the location of migration. There are reports in the literature describing migration of IVC filters but few report fracture of the fllter with subsequent migration and the risk of this complication is unclear. In this report we discuss the risk of fllter fracture and outcomes related to this complication.

Case Report

This is a case report of a 20-year-old man with a history of ulcerative colitis (UC) diagnosed with a DVT in the right popliteal vein as well as a catheter associated left basilic vein thrombosis during a hospitalization for an exacerbation ofUC. He was initially treated with dalteparin, at a dose of200 units/ kg/day subcutaneously for one month followed by a reduced dose of 175 units/kg/day, consistent with standard practice.

Two months after his initial episodes ofVTE, he developed

symptomatic extension of the initial right popliteal DVT into the right superficial femoral vein and he was admitted to hospital for intravenous unfractionated heparin then transitioned back to dalteparin 200 units/kg/day prior to

discharge from the hospital. A follow up ultrasound a month later revealed improvement of the clot and at the patient's request, dalteparin was discontinued without the initiation of warfarin after a total duration of anticoagulation of six months.

He presented a month later with new right leg pain and edema and an ultrasound revealed recurrent DVT in the right external iliac vein. At the same time he had a UC related lower gastrointestinal bleed with a hemoglobin of 58g/L. His UC was refractory to previous medical treatment and given the ongoing bleeding and transfusion requirements, he required a laparoscopic subtotal colectomy and end ileostomy. Given his history of recent extensive and recurrent DVT, as well as his high perioperative bleeding risk, a Gunther-Tulip IVC filter was placed pre-operatively. Systemic anticoagulation was resumed post-operatively with IV unfractionated heparin but despite this he developed an asymptomatic 4.1 x 1.5 em thrombus above the level of the filter, found incidentally on a follow-up ultrasound two weeks after the filter placement. He was transitioned to dalteparin adjusted according to peak anti­ factor Xa levels (targetl.2 ± 0.43).

After resolution of the thrombus on follow-up imaging, removal of the filter was planned, two months after its insertion. At the time of removal, the filter was found to be tilted with the apex imbedded in the posterior wall of the IVC and one of the legs resting in the right renal vein. Initial attempts at removal were unsuccessful. Imaging at the time of a second attempt to remove the filter 30 days later, showed that one of the legs had fractured and migrated to the left distal pulmonary artery

Figure 1. Images of chest revealing filter leg (arrow) in left pulmonary artery prior to and following remova.l


Ca n adia n Journal of Ge n eral I n terna l Medic ine Volume 11. Issue 3, 2016 37

Inferior Vena Cava Filter Fracture and Migration

-----------------------------------------------------------------------------------------------------

(Figure 1). The patient was asymptomatic and was maintained on dalteparin. A week later, the filter leg was removed using an endovascular approach and bronchoscopy forceps through an

8 French guiding sheath without complication.

Discussion

This is a case report that describes a fracture of a fllter leg and subsequent migration to the pulmonary artery. Filter fracture and migration is a recognized complication ofNC filters and the incidence reported in the literature varies with the type of

device and duration of placement (Table 1) with estimates of

severe consequences. Nicholson et al. reported a case of sudden death from fllter fracture and migration.4 Kumar described a case where a broken arm of an IVC fllter migrated to the heart, penetrated the right ventricle and required cardiac surgery.15

Hull reported a case of a patient presenting with non-sustained ventricular tachycardia after filter arm perforation, fracture and migration to the right ventricle.16 Several cases of fllter fracture with migration leading to cardiac tamponade have also been reported.4.l7·30 Other areas of fracture and migration include perforation of the abdominal aorta, duodenal perforation and

migration to the kidneyY·23 A case of fracture and migration of

1% to 29%.12'

Most studies commenting on rates of fllter

an OptEase filter to the pulmonary vasculature during retrieval

15

fracture are retrospective reviews with fracture and migration discovered on follow-up imaging. Tam et al, retrospectively, evaluated images of 363 patients with the Bard Recovery Filter and reported 26 fracture events (7%) at a median follow-up of

18.4 months.2 Similar rates of fllter fracture were reported by Dinglasan et al who identified 15 fractures out of 148 (10%) cases of Bard Recovery, G2 and Celect filters reviewed.3 Only one report is in the literature commenting on fracture rates for the Gunther-Tulip filter and in that retrospective review of 105 cases, no fllter fracture or migration occurred.11

Although less common than other filter complications, filter fracture and subsequent migration can have particularly

Table 1. Inferior vena cava fracture rates

that required surgical removal was recently reported.24

The use of IVC filters should be considered only in patients with acute VTE who have a contraindication to anticoagulant therapy. Although filter fracture and migration is rare, it is a recognized and potentially serious iatrogenic complication. Removal of IVC filters is potentially fraught with hazard and it is important to inspect the filter to ensure it has been removed intact.

Acknowledgements: Thank you to Dr. Mehran Midia for providing the images.

Study

Filter Type

Median

Follow-Up (mo)

Number of

Filters

Number of

Fractures

Dinglasan et al.

G2,Recovery,Celect

13.5

148

15 (10%)

Vijay et al.

Recovery,G2,G2 Express

23.1

548

63(12%)

Hull et al.

Recovery

30.0

14

4(29%)

Tam et al.

Recovery

18.4

363

26 (7%)

Nicholson et al.

Recovery,G2

1.7,0.8

80

13 (16%)

lynch et al.

G2

6.0

174

6 (3%)

Zhu et al.

G2

4.3

139

2 (1%)

Sangwaiya et al.

Celect

2.3

18

1 (6%)

Johnson et al.

Option

6.0

100

0 (0%)

Oliva et al.

Recovery,G2

1.8

51

0 (0%)

Kalva et al.

TrapEase

6.3

270

8 (3%)

Kalva et al.

Recovery

2.7

40

3 (8%)

Rosenthal et al.

Gunther-Tulip

8.7

105

0

38 V o lume 1 1, Issue 3, 2016 Canad i an Journal o f Genera l Interna l Medicine

Lepic et al.

References

1. Hull JE, Robertson SW. Bard Recovery Filter: evaluation and management of vena cava limb perforation, fracture, and migration.J Vase Interv Radio!
2009;20:52 0.
2. Tam, MD et al. Fracture and distant migration of the Bard Recovery filter: a retrospective review of 363 implantations for potentially life-threatening complications. J Vase Interv Radio!2012;23(2):199-205.
3. Dinglasan LA, et al. Removal of fractured inferior vena cava filters:
feasibility and outcomes. J Vase Interv Radio!. 2012;23(2):181-7
4. Nicholson W et al. Prevalence of fracture and fragment embolization of Bard retrievable vena cava filters and clinical implications including cardiac perforation and tamponade. Arch Intern Med 2010;170(20):1827-31.
5. Vijay K. Fractured Bard Recovery, G2 and G2 Express inferior vena cava
filters: incidence,clinical conse.quences, and outcomes of removal attempts. J Vase Interv Radiol2012;23(2):188-94.
6. Lynch FC, Kekulawela S.Removal of the G2 filter: differences between implantation times greater and less than 180 Days. J Vase Interv Radio!
2009;20:1200-09.
7. Oliva, VL et al. Recovery G2 inferior vena cava filter: technical success and safety of retrieval.J Vase Interv Radiol 2008;19(6):884-9.
8. Zhu X et al. Retrievability and device-related complications of the G2
filter:a retrospective study of 139 filter retrievals. J Vasc lnterv Radio!
2011;22(6):806-12.
9. Kalva SP et al. Recovery vena cava filter:experience in 96 patients.
Cardiovasc Intervent Radiol2006;29(4):559-64.

10. Kalva SP. TrapEase vena cava filter: experience in 751 patients.J Endovasc

Ther 2006;13(3):365-72.

11. Rosenthal D et al. Retrievability of the GUnther Tulip vena cava filter after dwell times longer than 180 days in patients with multiple trauma.

J EndovascTher 2007:14(3}:406-10.
13. Johnson MS et al. The safety and effectiveness of the retrievable option inferior vena cava filter:a United States prospective multicenter clinical study. J Vase Interv Radiol2010;21(8):1173-84.
14. Sangwaiva MJ et a!. Safety and effectiveness of the Celect inferior vena cava filter:preliminary results. J Vase Interv Radiol2009;20(9):1188-92.
15. Kumar SP, Mahtabifard A, Young JN. Fractured inferior vena cava filter
strut presenting as a penetrating foreign body in the right ventricle: report of a case. } Card Surg 2008;23(4):378-81
16. Hull )E, Han J, Giessel GM. Retrieval of the Recovery filter after arm perforation, fracture, and migration to the right ventricle. J Vasc Interv Radiol 2008;19(7}:1107-11.
17. Rogers, NA eta!. Fracture and embolization of an inferior vena cava filter strut leading to cardiac tamponade. Circulation. 2009;119(18):2535-6
18. Saeed, I, Garcia, M, McNicholas K. Right ventricular migration of a Recovery IVC filter's fractured wire with subsequent pericardia! tamponade. Cardiovasc Intervent Radio!. 2006;129(4):685-6
19. Chandra, PA et al. Cardiac tamponade caused by fracture and migration of inferior vena cava filter. South Med J. 2008;101 (II):1163-4
20. Kalavakunta, JK, Thomas, CS, Gupta V. A needle through the heart:
rare complication of inferior vena caval filters. J Invasive Cardiol.
2009;21(11):E221-3
21. Dabbagh, A. Late complication of a Greenfield filter associating caudal migration and perforation of the abdominal aorta by a ruptured strut.
J VascSurg. 1995;22(2):182-7
22. Taheri SA et al. A complication of the Greenfield fllter: fracture and distal migration of two struts- a case report. } Vase Surg. 1992;16(1):96-9
23. Lamotte, V et al. Intrarenal migration of two fractured struts of a vena cava filter. Ann Radio!. 1993;36(4}:315-8
24. Hill, E et al. Vena cava filter fracture with migration to the pulmonary
artery. Ann Thorac Surg. 2003;95:342-5

Canadia n Journa l of General Interna l Medic ine Volume t 1, Issue 3, 2016 39